# Principles of ultrasonic imaging
Ultrasonic imaging is a portable technique with diverse applications that relies on the fundamental physics of sound wave propagation, reflection, and detection to generate images.
### 1.1 Nature of sound waves
Sound is a longitudinal wave, also known as a pressure wave, which requires a medium for propagation. It cannot travel through a vacuum, such as outer space. The characteristics of sound waves are described by:
* **Period ($T$)**: The time it takes for one complete cycle of the wave.
* **Frequency ($f$)**: The number of cycles per unit time, measured in Hertz (Hz). It is the inverse of the period:
$$f = \frac{1}{T}$$
* **Wavelength ($\lambda$)**: The spatial extent of one complete cycle of the wave, measured in distance units.
* **Speed ($c$)**: The rate at which the wave propagates through the medium.
These properties are interrelated by the formula:
$$c = f \lambda$$
Waves can be mathematically represented by sinusoidal functions. Waves with the same wavelength and amplitude can differ by a phase shift, which represents a displacement in time or space and is measured in radians.
Interference occurs when waves combine. **Constructive interference** amplifies the wave, while **destructive interference** cancels it out. Noise-canceling headphones utilize the principle of destructive interference.
The speed of sound is dependent on the properties of the propagation medium, specifically its density ($\rho$) and its elastic properties. For instance, the speed of sound in the human body is higher than in air.
### 1.2 Pressure and intensity
Sound energy causes particle displacement and variations in local pressure within a medium.
* **Pressure amplitude ($P$)**: Represents the maximum deviation from the average pressure in the medium in the absence of a sound wave. The SI unit for pressure is the Pascal (Pa), defined as 1 Newton per square meter (N/m²). Diagnostic ultrasound waves typically have peak pressure levels around 1.5 MPa, which is approximately 15 times atmospheric pressure.
* **Acoustic intensity ($I$)**: Defined as the average energy transmitted per unit time per unit area, perpendicular to the direction of wave propagation. It is related to the pressure amplitude by:
$$I \propto P^2$$
Acoustic intensity in medical diagnostic ultrasound is measured in milliwatts per square centimeter (mW/cm²). Regulatory bodies like the FDA limit these intensities to 750 mW/cm² for human medical use.
The decibel (dB) scale is used to express intensity levels logarithmically. A 10 dB change corresponds to a tenfold increase in intensity, and a 20 dB change corresponds to a hundredfold increase.
> **Tip:** Understanding the relationship between intensity and pressure is crucial for comprehending how ultrasound energy interacts with tissues.
### 1.3 Interactions between ultrasound waves and matter
The behavior of ultrasonic waves interacting with matter is governed by the acoustic properties of the materials.
* **Acoustic impedance ($Z$)**: A material property defined as the product of its density ($\rho$) and the speed of sound within it ($c$):
$$Z = \rho c$$
The SI unit for acoustic impedance is the rayl, which is equivalent to kg/(m²·s). A higher acoustic impedance indicates greater resistance to the propagation of sound waves. For example, air has a very low acoustic impedance, while bone has a high acoustic impedance.
When an ultrasound wave encounters an interface between two media with different acoustic impedances, reflection and transmission occur. The amplitude of the reflected wave depends on the difference in acoustic impedances.
* **Reflection**: Occurs at interfaces. If the interface is smooth relative to the wavelength of the ultrasound, specular reflection occurs. If the interface is rough, diffuse scattering occurs.
* **Refraction**: Bending of the ultrasound beam as it passes from one medium to another with a different speed of sound. The angle of refraction depends on the angle of incidence and the ratio of the speeds of sound in the two media, as described by Snell's Law. For non-perpendicular incidence, the angle of reflection equals the angle of incidence ($\theta_i = \theta_r$).
**Scattering** happens when ultrasound waves interact with objects or interfaces that are comparable in size to the wavelength or smaller. The amplitude of the echo signal is influenced by:
* The number and size of scatterers per unit volume.
* The differences in acoustic impedance at the interfaces of the scatterers.
* The ultrasound frequency.
Higher frequency ultrasound has shorter wavelengths. While this can improve resolution, it also leads to increased attenuation and reduced penetration depth.
> **Example:** The interface between soft tissue and bone has a large difference in acoustic impedance, resulting in significant reflection and scattering of ultrasound. This is why bone is a poor medium for ultrasound imaging, as it hinders the passage of sound to deeper structures.
**Attenuation** is the reduction in ultrasound intensity as it propagates through a medium. It is caused by absorption (conversion to heat) and scattering. Materials like bone exhibit high attenuation.
### 1.4 The transducer: emission and detection of ultrasound
The **transducer** is the key component of an ultrasound system responsible for both generating and receiving ultrasound waves. It converts electrical energy into mechanical energy (ultrasound) and vice versa.
* **Piezoelectric crystal**: The core of the transducer is a piezoelectric crystal. When an electric field is applied across this crystal, it vibrates and produces mechanical waves (ultrasound). Conversely, when mechanical waves strike the crystal, they cause it to deform, generating an electrical voltage.
The transducer typically includes:
* **Backing block**: Absorbs excess vibrations to limit the duration of the emitted pulse and improve image quality.
* **Matching layer**: Placed between the piezoelectric crystal and the object being imaged to minimize reflections at the interface due to acoustic impedance mismatches, thereby improving the transmission of ultrasound into the body.
Transducers can be composed of a single piezoelectric element or an array of elements.
* **Linear array transducers**: Elements are arranged in a straight line and are activated sequentially or in groups to steer the beam linearly.
* **Phased array transducers**: Elements are arranged in a way that allows for electronic steering and focusing of the ultrasound beam by introducing time delays in the activation of individual elements. This is achieved through **beamformer electronics**.
The ultrasound beam has distinct regions:
* **Near field (Fresnel zone)**: The region close to the transducer where the beam is complex and can be focused.
* **Far field (Fraunhofer zone)**: The region further from the transducer where the beam diverges and is less focused. The best lateral resolution is typically achieved in the focal zone, at the transition between the near and far fields.
### 1.5 Image formation and display modes
Ultrasound imaging creates images based on the timing and amplitude of returning echoes.
* **Pulsed ultrasound**: The transducer emits short bursts (pulses) of ultrasound and then listens for the returning echoes.
* **Pulse repetition frequency (PRF)**: The number of pulses emitted per second, typically in the range of 2 to 7 kHz. It is determined by the desired imaging depth, as sufficient time must be allowed for echoes from distant structures to return. The maximum depth that can be imaged is related to the speed of sound and the pulse repetition period (PRP), where $D_{max} = \frac{c \cdot PRP}{2}$.
* **Pulse repetition period (PRP)**: The time between the start of successive pulses, equal to the inverse of the PRF ($PRP = 1/PRF$).
* **Pulse duration**: The time for which a single pulse is transmitted, determined by the number of cycles in the pulse and the transducer frequency.
* **Duty cycle**: The fraction of time the transducer is actively transmitting, calculated as $Pulse Duration / PRP$. For real-time imaging, this is typically very low (0.2-0.4%).
Ultrasound systems operate in either a **pulsed mode** (high voltages, e.g., ~150 V) for transmission or a **receive mode** (low voltages, e.g., ~1 V to 2 µV) for listening to echoes.
Images are formed from a series of **A-lines**, which represent the amplitude of received echoes along a specific direction as a function of depth. Different display modes are used to visualize this information:
* **A-mode (Amplitude mode)**: Displays the amplitude of the echo signal as a function of depth.
* **B-mode (Brightness mode)**: Converts the amplitude of the echo signal into a brightness level on a 2D display, creating a cross-sectional image. Multiple A-lines are compiled to form a B-mode image.
* **M-mode (Motion mode)**: Displays a single A-line over time, allowing for the visualization and quantification of motion of structures, such as heart valves.
**Real-time imaging** involves acquiring and displaying multiple 2D images (frames) per second. The frame rate depends on several factors, including:
* **Field of view (FOV)**: The sector angle of the image.
* **Imaging depth ($D$)**: The maximum depth visualized.
* **Line density (LD)**: The number of A-lines used to construct the image, which is related to the FOV and the number of lines ($N$).
* **Time to acquire one frame ($T_{frame}$)**: The total time to capture all A-lines for a single image, often calculated as $N \times T_{line}$, where $T_{line}$ is the time to acquire one A-line.
Increasing the frame rate can be achieved by reducing the imaging depth, decreasing the number of A-lines, or narrowing the FOV, often involving a trade-off with image quality or resolution.
### 1.6 Spatial resolution
Spatial resolution refers to the ability to distinguish between two closely spaced objects. In ultrasound imaging, it is typically described in three dimensions:
* **Axial resolution**: The ability to resolve objects along the direction of the ultrasound beam. It is primarily determined by the spatial pulse length (SPL) of the transmitted pulse. Shorter pulses (higher frequency, shorter pulse duration) lead to better axial resolution.
* **Lateral resolution**: The ability to resolve objects perpendicular to the ultrasound beam within the plane of the beam. It is determined by the beam diameter and is generally worse than axial resolution, especially in the far field where the beam diverges.
* **Elevation resolution**: The ability to resolve objects perpendicular to both the beam direction and the image plane. It is dependent on the height of the transducer elements.
Higher ultrasound frequencies (shorter wavelengths) generally result in better spatial resolution but are associated with reduced penetration depth due to increased attenuation. The choice of frequency depends on the specific application and the depth of the structures being imaged.
> **Tip:** For deeper structures, lower frequencies (e.g., 3.5-5 MHz) are used to achieve adequate penetration, while for superficial structures, higher frequencies (e.g., 7.5-10 MHz) provide better resolution. Very high frequencies (30-40 MHz) can be used for intravascular imaging where penetration is not a significant issue.
### 1.7 Doppler ultrasound
Doppler echocardiography utilizes the Doppler effect to assess blood flow. When ultrasound waves reflect off moving structures, such as red blood cells, the frequency of the returning echo is shifted relative to the transmitted frequency. This **Doppler frequency shift** is proportional to the velocity of the moving object and the cosine of the angle between the ultrasound beam and the direction of motion.
The Doppler equation is often expressed as:
$$f_d = \frac{2 f_c v \cos \theta}{c}$$
where:
* $f_d$ is the Doppler frequency shift.
* $f_c$ is the transmitted (carrier) frequency.
* $v$ is the velocity of the moving object.
* $\theta$ is the angle between the ultrasound beam and the direction of motion.
* $c$ is the speed of sound.
The factor of 2 arises because the Doppler shift occurs twice: once when the wave reflects off the moving target and again as the reflected wave is detected by the transducer.
> **Note:** Large angles ($\theta > 60^\circ$) between the transducer and the flow direction lead to small Doppler shifts, making velocity estimation highly sensitive to angle errors.
The **Doppler spectrum** displays the distribution of Doppler frequencies (and thus velocities) over time. Normal blood flow has a characteristic spectral waveform related to the hemodynamics of the vascular system. Abnormalities like turbulent flow result in altered spectra.
Quantitative parameters derived from Doppler measurements include:
* **Pulsatility Index (PI)**: $\frac{(V_{max} - V_{min})}{V_{mean}}$
* **Resistive Index (RI)**: $\frac{(V_{max} - V_{min})}{V_{max}}$
These indices help quantify the pulsatility and resistance of blood flow in arteries.
### 1.8 Other applications and considerations
* **Distance measurements**: Ultrasound can accurately measure distances, areas, and volumes by timing the transit of the ultrasound pulse to and from anatomical structures. The speed of sound in soft tissue is approximately 1540 m/s, allowing for precise calibration.
* **Contrast agents**: Microbubbles (1-10 µm in diameter) containing gas are used as contrast agents to enhance echogenicity. They are small relative to the ultrasound wavelength and act as strong scatterers due to the large difference in acoustic impedance between the microbubble and the surrounding tissue.
* **Biological effects**: Ultrasound can have thermal and mechanical effects on tissues.
* **Thermal effects**: Absorption of ultrasound energy leads to tissue heating. The **Thermal Index (TI)** quantifies the potential for thermal effects.
* **Mechanical effects**: Primarily cavitation (formation and collapse of microbubbles), which can cause tissue damage. The **Mechanical Index (MI)** is a measure of the likelihood of cavitation.
At diagnostic imaging intensity levels and durations, ultrasound is considered a very safe imaging modality.
**Advantages of ultrasound imaging** include its non-invasiveness, portability, relatively low cost, availability, lack of ionizing radiation, and real-time imaging capability.
**Disadvantages** include its strong dependence on operator skill for image acquisition and interpretation, limitations in imaging through air and bone, difficulty imaging deep structures, and generally lower resolution and contrast compared to CT and MRI.
---
# Ultrasound equipment and data acquisition
Ultrasound imaging is a widely used, portable imaging technique that relies on the reflection of sound waves to generate images.
### 2.1 Fundamentals of ultrasound imaging
#### 2.1.1 Sound propagation
Sound is a longitudinal wave, also known as a pressure wave, that requires a medium to propagate. Its characteristics are defined by:
* **Period ($T$)**: The time for one complete cycle of a wave.
* **Wavelength ($\lambda$)**: The spatial distance over which a wave's shape repeats.
* **Frequency ($f$)**: The number of cycles per unit time, measured in Hertz (Hz). It is inversely related to the period: $f = 1/T$.
* **Speed ($c$)**: The distance traveled per unit time. The relationship between speed, wavelength, and frequency is given by: $c = \lambda f$.
* **Amplitude**: The maximum displacement or pressure variation from the equilibrium value.
* **Phase difference**: The temporal or spatial shift between waves with the same wavelength and amplitude, often expressed in radians.
Interference occurs when waves overlap. Constructive interference leads to an increase in amplitude, while destructive interference leads to a decrease.
The speed of sound in a medium depends on its properties, primarily its density ($\rho$) and its bulk modulus (which relates to compressibility). This relationship can be expressed as: $c = \sqrt{\frac{K}{\rho}}$, where $K$ is the bulk modulus.
#### 2.1.2 Pressure and intensity
* **Pressure amplitude ($P$)**: The maximum variation in pressure from the ambient pressure caused by the sound wave. Diagnostic ultrasound peak pressure levels are around $1.5$ MPa, approximately $15$ times atmospheric pressure.
* **Acoustic intensity ($I$)**: The average power per unit area, perpendicular to the direction of wave propagation. It is related to the pressure amplitude by: $I \propto P^2$. Medical diagnostic ultrasound intensity levels are typically in the milliwatts per square centimeter ($mW/cm^2$) range, with FDA limits set at $750$ $mW/cm^2$.
A $10$-fold increase in intensity corresponds to a $10$ dB change on the decibel scale, and a $100$-fold increase corresponds to a $20$ dB change.
#### 2.1.3 Acoustic impedance
Acoustic impedance ($Z$) is a material property that describes its resistance to the propagation of sound waves. It is defined as the product of the material's density ($\rho$) and the speed of sound ($c$) within it:
$$Z = \rho c$$
The SI unit for acoustic impedance is rayl, which is equivalent to $1$ $kg/(m^2 \cdot s)$.
**Tip:** Acoustic impedance differences at tissue interfaces are crucial for generating ultrasound echoes. Air has a very low acoustic impedance, while bone has a high acoustic impedance.
#### 2.1.4 Interactions with matter
When an ultrasound wave encounters an interface between two media with different acoustic impedances, some of the wave is reflected, and some is transmitted. The amplitude of the reflected wave depends on the difference in acoustic impedances.
* **Reflection and refraction**: If the angle of incidence is not perpendicular, the wave reflects at an equal angle, and refraction (bending of the wave) occurs according to Snell's Law, dependent on the change in sound speed.
* **Scattering**: Ultrasound waves can be scattered by objects or interfaces smaller than or comparable to the wavelength.
* **Specular reflectors**: These are smooth surfaces relative to the ultrasound wavelength, causing reflection in a specific direction.
* **Non-specular reflectors**: These are rough surfaces or small objects that scatter the ultrasound energy in multiple directions.
Higher frequency ultrasound (shorter wavelength) interacts with smaller structures, leading to more diffuse scattering and a smaller fraction of the incident intensity returning to the transducer.
* **Attenuation**: As ultrasound propagates through tissue, its intensity decreases due to absorption (conversion to heat) and scattering. Tissues like bone have high attenuation, making imaging through them difficult. Air also causes significant attenuation and reflection. Viscosity, a measure of a fluid's resistance to deformation, also contributes to attenuation.
The fundamental principle of ultrasound imaging lies in the differences in acoustic impedance and sound speed between tissues, which cause reflections (echoes) that are detected by the transducer.
### 2.2 Ultrasound equipment
#### 2.2.1 Transducer (converter)
The transducer is the core component of an ultrasound system, responsible for both emitting and receiving ultrasound waves. It utilizes the piezoelectric effect:
* **Emission**: Electrical energy is converted into mechanical energy, causing the piezoelectric crystal to vibrate and generate sound waves.
* **Reception**: Incoming mechanical vibrations from returning ultrasound waves are converted back into electrical energy, which can be processed.
The piezoelectric material, commonly PZT (lead zirconate titanate) or PVDF (polyvinylidene fluoride), deforms under an electric field and induces an electric field when deformed.
A transducer typically includes:
* **Piezoelectric crystal**: The active element.
* **Backing block**: Damps excessive vibrations.
* **Matching layer**: Reduces impedance mismatch between the crystal and the tissue, improving transmission and reception.
#### 2.2.2 Transducer arrays
Transducers can consist of multiple piezoelectric elements arranged in arrays, allowing for more sophisticated beam control.
* **Linear array**: Elements are arranged in a straight line. Activating elements sequentially or in groups can steer the beam linearly across the field of view.
* **Phased array**: Elements are arranged in a specific pattern. By introducing small time delays (phasing) in the activation of individual elements, the ultrasound beam can be steered electronically in different directions. This allows for sector scanning and focusing of the beam.
The **beam former** electronics control the activation of array elements to shape and steer the ultrasound beam. The beam has a **near field** (Fresnel zone) and a **far field** (Fraunhofer zone), with a focal zone where the beam diameter is smallest.
#### 2.2.3 Pulse modes and real-time imaging
Ultrasound data is acquired by transmitting short pulses of ultrasound and listening for the returning echoes.
* **Pulsed mode**: High voltages (around $150$ V) are applied to excite the transducer.
* **Listening mode**: Low voltages (around $1$ V down to $2$ $\mu$V) are used to detect weak returning echo signals.
Key parameters in pulsed ultrasound:
* **Pulse Repetition Frequency (PRF)**: The number of pulses transmitted per second. Typically $2$ to $7$ kHz. The PRF is determined by the desired imaging depth.
$$PRF = \frac{1}{PRP}$$
* **Pulse Repetition Period (PRP)**: The time interval between the start of consecutive pulses. It is the inverse of the PRF.
* **Pulse duration**: The time during which a single pulse is emitted. It is determined by the number of cycles in the pulse and the transducer frequency.
* **Duty cycle**: The fraction of time the transducer is actively transmitting a pulse. It is calculated as:
$$Duty \ Cycle = \frac{Pulse \ Duration}{PRP}$$
For real-time imaging, the duty cycle is typically very low, around $0.2\%$ to $0.4\%$.
**Tip:** It's crucial to distinguish between the ultrasound wave frequency (typically in MHz) and the PRF (typically in kHz). Similarly, the period of an ultrasound wave (in $\mu$s) is different from the PRP (in ms).
**Maximum imaging depth ($D_{max}$)**: The maximum depth from which echoes can be received before the next pulse is transmitted is limited by the PRP. This depth is given by:
$$D_{max} = \frac{c \times PRP}{2}$$
The division by $2$ accounts for the round trip of the sound pulse.
#### 2.2.4 Spatial resolution
Spatial resolution refers to the ability to distinguish between two closely spaced objects. There are three main types:
* **Axial resolution**: Resolution along the direction of the ultrasound beam. It is primarily determined by the pulse length. Shorter pulses lead to better axial resolution.
* **Lateral resolution**: Resolution perpendicular to the ultrasound beam but within the plane of the beam. It is determined by the beam diameter and is depth-dependent, being best at the focal zone.
* **Elevational resolution**: Resolution perpendicular to the imaging plane. It is influenced by the height of the transducer elements.
Higher ultrasound frequencies (shorter wavelengths) generally provide better resolution but have reduced penetration depth due to increased attenuation.
**Tip:** The choice of ultrasound frequency depends on the application:
* Deeper structures (e.g., abdominal imaging): $3.5$ to $5$ MHz.
* Superficial structures (e.g., thyroid, breast): $7.5$ to $10$ MHz.
* Intravascular imaging: $30$ to $40$ MHz for very high resolution.
### 2.3 Data acquisition
Data acquisition in ultrasound involves transmitting pulses and processing the returning echoes to form an image.
#### 2.3.1 Echo modes
Different modes of displaying ultrasound data are used:
* **A-mode (Amplitude mode)**: Displays the amplitude of the returning echo signal as a function of depth along a single line (scan line). It shows the strength and depth of reflectors.
* **B-mode (Brightness mode)**: Converts the echo amplitude into a brightness level on a display. A B-mode image is formed by assembling multiple A-lines from different scan directions, creating a 2D cross-sectional image.
* **M-mode (Motion mode)**: Displays the movement of structures over time. A single A-line is repeatedly sampled, and these samples are displayed side-by-side as a function of time, creating a trace of motion. This is useful for evaluating moving structures like heart valves.
#### 2.3.2 Real-time imaging
Real-time ultrasound imaging allows for the visualization of dynamic processes. A 2D image, known as a **frame**, is constructed from a series of A-lines.
* **Field of View (FOV)**: The angular extent of the sector scanned.
* **Imaging Depth ($D$)**: The maximum depth being visualized.
* **Frame Rate**: The number of frames displayed per second. It is inversely related to the time required to acquire one frame ($T_{frame}$).
$$T_{frame} = N \times T_{line}(D)$$
where $N$ is the number of A-lines and $T_{line}(D)$ is the time to acquire one line to depth $D$.
* **Line Density (LD)**: The number of A-lines per unit angle within the FOV.
**Tip:** Increasing the frame rate can be achieved by reducing the imaging depth ($D$), decreasing the number of A-lines ($N$), or reducing the FOV. This involves trade-offs between temporal resolution, spatial resolution, and image quality.
#### 2.3.3 Doppler echocardiography
Doppler echocardiography utilizes the Doppler effect to measure blood flow velocity. The Doppler frequency shift ($\Delta f$) is the difference between the transmitted and received frequencies and is proportional to the velocity of the moving object and the cosine of the angle ($\theta$) between the ultrasound beam and the direction of motion:
$$\Delta f = \frac{2 f_0 v \cos \theta}{c}$$
where $f_0$ is the transmitted frequency, $v$ is the velocity of the object, and $c$ is the speed of sound.
**Tip:** Large angles (greater than $60^\circ$) lead to small Doppler shifts, making velocity estimation highly sensitive to errors in angle measurement.
Doppler analysis can also provide information about flow patterns:
* **Doppler Spectrum**: Displays the distribution of velocities within the sample volume as a function of time. Deviations from normal spectral waveforms can indicate disturbed or turbulent flow, correlating with disease.
* **Pulsatility Index (PI)**: A measure of the difference between peak systolic and end-diastolic velocities relative to the mean velocity:
$$PI = \frac{V_{max} - V_{min}}{V_{mean}}$$
* **Resistive Index (RI)**: A measure of resistance in the vasculature, calculated as:
$$RI = \frac{V_{max} - V_{min}}{V_{max}}$$
#### 2.3.4 Contrast agents
Ultrasound contrast agents are microbubbles (typically $1$ to $10$ $\mu$m in diameter) containing gases like air or nitrogen, encapsulated in biocompatible materials. They have a significantly different acoustic impedance than surrounding tissue, leading to strong reflections and enhanced contrast in ultrasound images. They are used to improve visualization of blood vessels and perfusion.
#### 2.3.5 Measurements and biological effects
* **Distance measurements**: Accurate distance measurements are possible by timing the pulse-echo travel time and knowing the speed of sound in tissue (approximately $1540 \pm 15$ m/s). This allows for measurements of length, area, and volume.
* **Biological effects**: Ultrasound can cause both thermal and mechanical effects in biological tissues.
* **Thermal effects**: Caused by the absorption of ultrasound energy and conversion to heat. The **Thermal Index (TI)** is a parameter indicating the potential for temperature elevation.
* **Mechanical effects**: Primarily **cavitation**, which involves the formation and collapse of microbubbles due to pressure variations. The **Mechanical Index (MI)** is a measure of the likelihood of cavitation.
At diagnostic imaging levels, ultrasound intensities and exposure times are generally kept low, making it a very safe imaging modality. Macroscopic damage (e.g., cell lysis) is observed at much higher intensities, often used in therapeutic applications like lithotripsy.
### 2.4 Advantages and disadvantages of ultrasound imaging
**Advantages:**
* Generally non-invasive.
* Widely available and relatively inexpensive.
* No ionizing radiation.
* Provides real-time imaging.
* Considered a very safe technique.
**Disadvantages:**
* Image acquisition and interpretation are highly operator-dependent.
* Air and bone impede sound wave propagation, limiting visibility of underlying structures.
* Imaging deep structures can be challenging.
* Resolution and contrast may be lower compared to CT or MRI.
---
# Advanced ultrasonic imaging techniques and applications
This section delves into specialized ultrasonic imaging methods that extend beyond basic B-mode imaging, focusing on their principles and applications for enhanced diagnostic capabilities.
### 3.1 Doppler echography
Doppler echography is a specialized technique that utilizes the Doppler effect to assess blood flow. The Doppler effect, analogous to the change in pitch of a siren as it approaches or recedes, describes the frequency shift observed when an ultrasound wave reflects off moving objects. In the context of medical imaging, this moving object is typically blood.
#### 3.1.1 Principles of Doppler echography
The Doppler frequency shift ($f_d$) is directly proportional to the velocity ($v$) of the moving object (e.g., blood), the frequency of the transmitted ultrasound wave ($f_0$), and the speed of sound in the medium ($c$). It is also dependent on the angle ($\theta$) between the ultrasound beam and the direction of motion. The relationship is given by:
$$f_d = \frac{2 f_0 v \cos(\theta)}{c}$$
The factor of two arises because the Doppler shift occurs twice: once when the wave reflects off the moving object, and again when the moving object acts as a moving reflector for the returning wave.
* **Frequency shift:** A positive frequency shift occurs when the reflector (blood) is moving towards the transducer, indicating an increased frequency. A negative shift occurs when the reflector is moving away, indicating a decreased frequency.
* **Angle dependency:** The $\cos(\theta)$ term highlights the critical importance of the angle between the ultrasound beam and the blood flow. At angles close to 0 degrees (beam parallel to flow), the cosine is close to 1, maximizing the detected Doppler shift. As the angle increases towards 90 degrees (beam perpendicular to flow), the cosine approaches 0, and the Doppler shift becomes minimal or undetectable. Large angles (greater than 60 degrees) result in small Doppler frequency shifts, leading to significant inaccuracies in velocity estimation due to even small errors in angle measurement.
#### 3.1.2 Doppler spectrum and quantification
Doppler ultrasound can display the distribution of velocities within the sample volume as a spectrum, showing the amplitude of different velocities over time.
* **Normal blood flow:** Characterized by a specific spectral waveform that reflects the hemodynamic properties of the blood vessel.
* **Abnormal blood flow:** Disturbed or turbulent blood flow, often associated with pathological conditions, results in altered and disordered spectra. Quantifying these spectral changes can help detect abnormalities and correlate them with disease processes.
Several parameters are used to quantify Doppler spectra:
* **Pulsatility index (PI):** Defined as the difference between the peak systolic and end-diastolic velocities, divided by the mean velocity.
$$PI = \frac{v_{\text{max}} - v_{\text{min}}}{\bar{v}}$$
PI provides a measure of the variation in blood flow velocity during the cardiac cycle.
* **Resistive index (RI):** Defined as the difference between the peak systolic and end-diastolic velocities, divided by the peak systolic velocity.
$$RI = \frac{v_{\text{max}} - v_{\text{min}}}{v_{\text{max}}}$$
RI is an indicator of the resistance to blood flow in an artery.
### 3.2 Measurements of distance, area, and volume
Ultrasonic imaging is routinely used for quantitative measurements of distances, areas, and volumes.
* **Distance measurement:** The time taken for an ultrasound pulse to travel to a structure and for its echo to return is precisely measured. Knowing the speed of sound in soft tissue (approximately $1540 \pm 15$ meters per second), the depth of the structure can be accurately calculated. The instrument's timing accuracy allows for easy calibration. This principle is applied in measurements such as fetal femur length and abdominal circumference.
* **Area and Volume calculation:** By tracing the boundaries of organs or lesions on B-mode images, their areas and subsequently their volumes can be calculated.
### 3.3 Contrast agents in ultrasonography
Contrast agents are substances injected intravenously to enhance the visibility of blood flow and vascular structures, thereby improving image quality and diagnostic confidence.
#### 3.3.1 Principles of ultrasound contrast agents
Ultrasound contrast agents typically consist of microbubbles, usually 1 to 10 micrometers in diameter. These microbubbles are small relative to the ultrasound wavelength.
* **Composition:** They are often filled with air, nitrogen, or other gas compounds and encapsulated in materials like human albumin or perfluorocarbons to ensure stability.
* **Mechanism of action:** The gas-filled microbubbles have significantly different acoustic impedances compared to surrounding tissues. This large impedance mismatch leads to strong reflection of ultrasound waves, generating intense echoes that enhance contrast in the image.
* **Behavior in the vasculature:** Contrast agents are designed to remain stable in the bloodstream for a sufficient duration, allowing them to reach and opacify specific vascular areas targeted by the examination.
* **Imaging phases:** The appearance of contrast agents can be observed in different phases:
* **Arterial phase:** Shows opacification of arteries.
* **Late phase:** Shows clearance of the contrast agent from tissues.
* **Without contrast:** A baseline image without the agent.
#### 3.3.2 Applications of contrast-enhanced ultrasound
Contrast-enhanced ultrasound (CEUS) has found applications in various fields, particularly in the characterization of liver lesions. For instance, CEUS can differentiate between a normal thrombus and a neoplastic thrombus (a tumor within a blood vessel) by observing the contrast enhancement patterns during different phases of contrast administration.
### 3.4 Biological effects of ultrasound
The interaction of ultrasound with biological tissues can lead to both thermal and mechanical effects.
#### 3.4.1 Thermal effects
Ultrasound energy is absorbed by tissues and converted into heat. The extent of heating depends on the rate of heat deposition and the tissue's ability to dissipate heat.
* **Thermal Index (TI):** A parameter that quantifies the potential for thermal effects. It is the ratio of the acoustic power produced by the transducer to the power required to raise the tissue temperature in the beam path by 1 degree Celsius. Higher TI values indicate a greater risk of thermal injury.
#### 3.4.2 Mechanical effects
Mechanical effects are primarily related to cavitation, which is the formation and collapse of microscopic gas bubbles in the ultrasound field.
* **Cavitation:** Negative pressure gradients in the ultrasound wave can lead to the formation of microbubbles. As pressure builds, these bubbles can collapse violently, potentially causing tissue damage.
* **Mechanical Index (MI):** A parameter that indicates the likelihood of cavitation. Higher MI values suggest a greater risk of mechanical bioeffects.
#### 3.4.3 Safety considerations
While high ultrasound intensities and prolonged exposure can cause macroscopic and microscopic damage (e.g., cell lysis, chromosome breakage), diagnostic ultrasound imaging typically operates at much lower intensity levels and for limited durations. This, combined with monitoring of TI and MI, makes diagnostic ultrasound a very safe imaging modality.
> **Tip:** Understanding the principles behind Doppler shifts, the role of angle, and the parameters for quantification (PI, RI) is crucial for interpreting flow patterns. Remember that higher frequencies provide better resolution but have reduced penetration depth.
---
# Biological effects and safety of ultrasound
This section outlines the potential biological effects of ultrasound, focusing on thermal and mechanical mechanisms, and discusses the safety considerations and comparative advantages and disadvantages of ultrasound in medical imaging.
### 4.1 Biological effects of ultrasound
Biological effects from ultrasound arise from its interaction with tissues, primarily through two mechanisms: thermal and mechanical.
#### 4.1.1 Thermal effects
When biological tissue absorbs ultrasound energy, this energy is converted into heat, leading to a temperature increase. The extent of this warming depends on the rate at which heat is deposited by the ultrasound beam and the efficiency with which that heat can be dissipated from the tissue.
The **Thermal Index (TI)** is a parameter used to quantify the potential for thermal effects. It is defined as the ratio of the acoustic power produced by the transducer to the power required to raise the temperature of the tissue within the ultrasound beam by one degree Celsius. A higher TI indicates a greater potential for tissue heating.
#### 4.1.2 Mechanical effects
Mechanical effects are primarily related to cavitation, which is the formation and collapse of microscopic gas bubbles within tissues. These bubbles can form due to pressure differences, particularly negative pressure variations in the ultrasound wave. When these bubbles collapse under increasing pressure, they can generate localized shock waves and high temperatures, potentially leading to tissue damage.
The **Mechanical Index (MI)** is a measure of the likelihood of cavitation occurring from an ultrasound beam. It is calculated based on parameters related to the acoustic pressure and the ultrasound frequency. A higher MI suggests a greater probability of cavitation.
#### 4.1.3 Observed biological damage
Biological effects have been demonstrated at high ultrasound intensity levels and with prolonged exposure durations.
* **Macroscopic damage** can include the physical disruption and rupture of cells. This principle is utilized in therapies like shock wave lithotripsy, where focused ultrasound is used to break down kidney stones.
* **Microscopic damage** can manifest as alterations at the cellular level, such as the breaking of chromosome strands or changes in the mitotic index (the rate of cell division).
It is important to note that for diagnostic imaging, ultrasound intensity levels are kept significantly lower than those at which biological effects have been observed, and the duration of examinations is typically limited. This ensures that diagnostic ultrasound is considered a very safe imaging modality.
### 4.2 Safety considerations and comparative analysis
The safety of ultrasound imaging is determined by the balance between the intensity of the ultrasound beam and the duration of exposure. Diagnostic ultrasound generally operates within parameters that minimize the risk of adverse biological effects.
#### 4.2.1 Advantages of ultrasound
Ultrasound imaging offers several significant advantages:
* **Non-invasive:** In most applications, ultrasound does not require surgical procedures or the insertion of instruments into the body.
* **Availability and cost-effectiveness:** Ultrasound machines are widely available and are generally less expensive than other advanced imaging modalities.
* **No ionizing radiation:** Unlike X-rays or CT scans, ultrasound does not use ionizing radiation, making it a safer option for pregnant women and children.
* **Real-time imaging:** Ultrasound provides immediate visualization of anatomical structures and physiological processes, allowing for dynamic assessment of movement and blood flow.
* **High safety profile:** As discussed, when used within established guidelines, ultrasound is considered a very safe diagnostic tool.
#### 4.2.2 Disadvantages of ultrasound
Despite its advantages, ultrasound also has limitations:
* **Operator-dependent:** The quality of image acquisition and interpretation relies heavily on the skill and experience of the sonographer or clinician.
* **Limitations with air and bone:** Ultrasound waves are significantly attenuated and reflected by air and bone. This prevents clear visualization of structures located behind these materials (e.g., lungs, brain through the adult skull).
* **Reduced penetration for deep structures:** Higher ultrasound frequencies, which offer better resolution, are more readily attenuated, making it challenging to image deep tissues effectively. Lower frequencies are used for deeper imaging, but at the cost of reduced resolution.
* **Lower resolution and contrast:** Compared to modalities like Computed Tomography (CT) and Magnetic Resonance Imaging (MRI), ultrasound may offer lower spatial resolution and contrast, particularly for certain types of tissue differentiation.
---
## Common mistakes to avoid
- Review all topics thoroughly before exams
- Pay attention to formulas and key definitions
- Practice with examples provided in each section
- Don't memorize without understanding the underlying concepts
Glossary
| Term | Definition |
|------|------------|
| Longitudinal wave | A wave in which the particles of the medium move parallel to the direction of energy transfer. Sound waves are a type of longitudinal wave, also known as a pressure wave. |
| Period | The time it takes for one complete cycle of a wave to occur. It is measured in units of time, such as seconds. |
| Wavelength | The spatial distance over which a wave's shape repeats. It is measured in units of distance, such as meters. |
| Frequency | The number of wave cycles that pass a given point per unit of time. It is measured in Hertz (Hz), where 1 Hz equals one cycle per second. |
| Destructive interference | A phenomenon where two waves combine to form a wave with a smaller amplitude than the individual waves, often resulting in cancellation. This principle is used in noise-canceling headphones. |
| Acoustic impedance | A physical property of a medium that describes its resistance to the propagation of sound waves. It is defined as the product of the medium's density ($\rho$) and the speed of sound (c) within it ($Z = \rho c$). |
| Specular reflector | A smooth surface between two different media that reflects sound waves in a specific direction, similar to how a mirror reflects light. The angle of incidence equals the angle of reflection. |
| Scattering | The redirection of waves in multiple directions due to interaction with objects or inhomogeneities in a medium. Acoustic scattering occurs when wave size is comparable to or smaller than the wavelength. |
| Attenuation | The gradual loss of wave intensity as it travels through a medium due to absorption and scattering. Tissues with high acoustic impedance, like bone, cause significant attenuation. |
| Transducer | A device that converts one form of energy into another. In ultrasound, a transducer converts electrical energy into mechanical (sound) energy to produce ultrasound waves and converts returning mechanical (echo) energy into electrical signals for detection. |
| Piezo-electric crystal | A material that exhibits the piezoelectric effect, meaning it deforms when subjected to an electric field and generates an electric charge when deformed. This property is fundamental to ultrasound transducer operation. |
| Phased array | A type of ultrasound transducer composed of multiple small piezoelectric elements that can be activated individually with precise timing. This allows for electronic steering and focusing of the ultrasound beam. |
| Beam former electronics | The electronic circuitry associated with a phased array transducer that controls the timing of the electrical pulses sent to each element. By introducing small time delays, it shapes and directs the ultrasound beam. |
| Focal zone | The region in the path of an ultrasound beam where the beam is most tightly focused, resulting in the best lateral resolution. This zone is determined by the transducer's design and the beam steering techniques. |
| Pulse repetition frequency (PRF) | The number of ultrasound pulses transmitted by the transducer per second. It is typically in the kilohertz (kHz) range and is related to the maximum imaging depth. |
| Pulse repetition period (PRP) | The time interval between the start of consecutive ultrasound pulses. It is the inverse of the pulse repetition frequency (PRP = 1/PRF). |
| Axial resolution | The ability of an ultrasound system to distinguish between two small structures along the direction of the ultrasound beam. It is primarily determined by the spatial pulse length. |
| Lateral resolution | The ability of an ultrasound system to distinguish between two small structures perpendicular to the direction of the ultrasound beam. It is primarily determined by the beam diameter. |
| Doppler effect | The change in frequency of a wave in relation to an observer who is moving relative to the wave source. In ultrasound, it is used to measure the velocity of moving reflectors, such as blood flow. |
| Doppler spectrum | A graphical representation of the frequency distribution of Doppler-shifted echoes, showing the range of velocities present in the sampled volume over time. It provides information about blood flow characteristics. |
| Pulsatility index (PI) | A quantitative measure of blood flow pulsatility, calculated as the difference between peak systolic and end-diastolic velocities divided by the mean velocity. It reflects the resistance in the arterial system. |
| Resistance index (RI) | A quantitative measure of vascular resistance, calculated as the difference between peak systolic and end-diastolic velocities divided by the peak systolic velocity. It is often used to assess peripheral arterial resistance. |
| Contrast agents | Injectable substances that temporarily alter the echogenicity of tissues or fluids, enhancing visualization of blood vessels or pathologies. They often contain microbubbles that interact strongly with ultrasound. |
| Thermal index (TI) | A parameter that estimates the potential for temperature elevation in tissues due to ultrasound absorption. It is a safety indicator for therapeutic and diagnostic ultrasound. |
| Mechanical index (MI) | A parameter that estimates the likelihood of cavitation (formation and collapse of bubbles) due to ultrasound. It is a safety indicator for diagnostic ultrasound, particularly at lower frequencies. |
| Cavitation | The formation and collapse of microscopic gas bubbles in a liquid medium when subjected to negative pressure fluctuations, such as those produced by ultrasound. This can lead to tissue damage. |